Product code

A40207

Name of the product

Basic Female CPR Training Manikin (Style 200)

Product description

MOUNTING AND INSTALLATION

Take the CPR dummy, display screen and manostat out of the package. Lay the manikin on its back and connect the display and manikin with the white cable, the manostat and the wrist of the CPR manikin with the gray cable. Then make the connection between the manostat and the power supply with the black cable. The CPR manikin is ready to use.

Turn on the power supply for use, and a female voice will say "Welcome to use our product, working start, please select the work way". There are three options:

  1. Drill Operation
  2. Single Operation
  3. Double Operation

After selecting the operating mode, the voice will speak again and say "Please select the work frequency". There are two options:

  1. 220 seconds
  2. 250 seconds (recommended)

After selecting the frequency, the voice will speak again and say “Please press the beginning key”. The work has begun. It should be completed within 250 seconds.

Airway Clearance: Put the CPR dummy on its back. The practitioner should close the nose using two fingers of the right hand and raise the chin by 70-90 degrees by placing the left hand behind the neck. In this way, the airway will be opened.

Mouth to Mouth Respiration:

  • If the air blowing capacity is less than 800 ml, the green light is on, the display shows 1 error.
  • If the air blowing capacity is between 800-1200 ml, the yellow light is on, the display shows 1 time correctly.
  • If the air blowing capacity is more than 1200 ml, the red light is on, the display shows 1 time error.

Correct Pressure to Chest: First of all, the right area to apply pressure on the chest should be determined. Then, the two hands are crossed and pressure is applied to the chest with both arms perpendicular to the mannequin. The depth of pressure to be applied is between 4-5 cm.

  • If the applied pressure is less than 4 cm, the green light is on, the display shows 1 time error.
  • If the applied pressure is between 4-5 cm, the yellow light will turn on, the display will show once correctly.
  • If the applied pressure is more than 5 cm, the red light is on, the display shows 1 time error.

DRILL OPERATION

This process trains students to grasp basic skills.

First, the airway should be opened. Then mouth-to-mouth breathing or chest pressure can be applied. The screen will show if these applications are correct. When the time reaches 120 seconds, the application time will expire and the screen will turn off.

Click the print button and keep the result as a reference for reporting. It will show the date, name, number, series, correct / incorrect pressure or blown number applied on the printed receipt.

SINGLE OPERATION

This procedure examines the only student performing first aid practice.

First, the airway should be opened. Give air twice and then apply 4 rounds within 250 seconds according to the standard of 30: 2 (30 times the chest pressure is applied, 2 times air is applied, and again 30 times the chest pressure is applied and 2 times air is given). This means the transaction was successful. Make sure that 2 times of air delivery and 30 times of chest pressure are all correct, wrong applications are excluded. If the process cannot be completed within 250 seconds, or if the process is not performed in accordance with the procedure, the transaction is unsuccessful. Press the Reset button for a single operation again. After a successful procedure, the music plays, the neck artery and heart start to beat, the pupils return to their normal size. These show that the human being is brought back to life.

Click the print button and keep the result as a reference for reporting. It will show the date, name, number, series, correct / incorrect pressure or blown number applied on the printed receipt.

DOUBLE OPERATION

This procedure examines two students performing first aid practice.

First, the airway should be opened. Give air twice and then apply 5 rounds within 250 seconds according to the standard of 30: 2 (30 times pressure is applied to the chest, 2 times air is applied, and again 30 times of pressure is applied to the chest and 2 times). This means the transaction was successful. Make sure that 2 times of air delivery and 30 times of chest pressure are all correct, wrong applications are excluded. If the process cannot be completed within 250 seconds, or if the process is not performed in accordance with the procedure, the transaction is unsuccessful. Press the Reset button for a double operation again. After a successful procedure, the music plays, the neck artery and heart start to beat, the pupils return to their normal size. These show that the human being is brought back to life.

Click the print button and keep the result as a reference for reporting. It will show the date, name, number, series, correct / incorrect pressure or blown number applied on the printed receipt.

CLEANING AND STORAGE CONDITIONS

Cover the mouth with gauze or wipe it with a tampon in case of mouth-to-mouth respiration against cross infections.

Her kullanımdan sonra, CPR modelinin parçaları (yüz, ağız, burun, göğüs, solunum borusu, hava valfi, vb.) deterjanla temizlenmeli ve %70 lik alkolle bir sonraki kullanım için dezenfekte edilmelidir.

When attaching the new lung bag, first loosen the screw, remove the old lung bag, attach the new one and tighten the nuts.

If the model will not be used for a long time, remove the batteries despite deterioration.

The model should be stored in a clean and dry place.

MATERIAL

The model is made of PVC plastic.

PACKING DIMENSIONS

94x38x58 cm

WEIGHT

  • Gross: 21 kg
  • Net: 19 kg